Document Type
Student Presentation
Presentation Date
4-21-2014
Faculty Sponsor
Lonny Ashworth
Abstract
Background: After reviewing the literature on both high flow and low flow Acapella devices, it was determined that most researchers evaluated only three frequency dial settings. Due to this gap, we chose to determine the expiratory resistance at each frequency dial setting for both high flow and low flow Acapella devices, as patient effort (amplitude) is increased on an electronic lung simulator. Hypothesis: 1) As the frequency dial setting is increased on the Acapella, the expiratory resistance will increase, and 2) as the patient effort (amplitude) is increased, the expiratory resistance will increase.
Methods: Each Acapella device was attached, separately, via a female-to-female adapter to the Hans Rudolph 1101 Electronic Lung Simulator. HR 1101 settings: Resistance 5 cm H2O/L/sec, Compliance 50 mL/cm H2O, Respiratory Rate 20/minute, Amplitude 10, 20, 30 and 40 cm H2O (to simulate patient effort), Percent Inhale 30 %, Targeted Volume 3000 mL, Load Effort Normal. Initially, amplitude was set at 10 cm H2O and the Acapella was set at the lowest setting. After allowing for stabilization, Peak Pressure and Peak Expiratory Flowrates were recorded for 20 consecutive breaths. The Acapella dial was then increased to the next setting; pressure and flowrates were recorded again as previously described. This process was continued until reaching the highest Acapella setting. Next, data was gathered at amplitudes of 20, 30 and 40 cm H2O, following the same procedure as stated for amplitude of 10 cm H2O. Resistance was calculated as (P1-P2)/Flow. P1 = averaged peak pressure for 20 breaths; P2 = 0 (ambient pressure); Flow = averaged Peak Expiratory Flowrate for 20 breaths.
Results: The expiratory resistance increased as the frequency dial setting was increased and the expiratory resistance increased as amplitude increased. At an amplitude of 20 cm H2O, the expiratory resistance increased from 25.83 to 47.02 cm H2O/L/sec on the blue Acapella and from 12.56 to 38.24 cm H2O/L/sec on the green Acapella device as the frequency dial setting was increased from 1-5 (Figure 1).
Conclusion: The expiratory resistance increased as the frequency dial setting increased from 1 to 5 on both Acapella devices. The expiratory resistance increased as the amplitude increased at 10, 20, 30 and 40 cm H2O, on both devices, confirming our hypothesis.
Recommended Citation
Wroblewski, Amanda and Hawkins, Scott, "Measured Expiratory Resistance of the Blue and Green Acapella Devices as Setting is Increased From 1–5; Amplitude 20, 30, 40" (2014). College of Health Sciences Presentations. 1.